TY - JOUR KW - Humans KW - Computational Biology KW - DNA Virus Infections KW - DNA Viruses KW - High-Throughput Nucleotide Sequencing KW - Metagenomics KW - Molecular Diagnostic Techniques KW - Sensitivity and Specificity KW - Transplantation AU - Meredith Carpenter AU - Susanna Tan AU - Thomas Watson AU - Rowena Bacher AU - Vaishnavi Nagesh AU - Alain Watts AU - Gordon Bentley AU - Jenna Weber AU - ChunHong Huang AU - Malaya Sahoo AU - Armin Hinterwirth AU - Thuy Doan AU - Theodore Carter AU - Queeny Dong AU - Stéphane Gourguechon AU - Eric Harness AU - Sean Kermes AU - Srihari Radhakrishnan AU - Gongbo Wang AU - Alejandro Quiroz-Zárate AU - Jesus Ching AU - Benjamin Pinsky AB -

Infections with DNA viruses are frequent causes of morbidity and mortality in transplant recipients. This study describes the analytical and clinical performance characteristics of the Arc Bio Galileo Pathogen Solution, an all-inclusive metagenomic next-generation sequencing (mNGS) reagent and bioinformatics pipeline that allows the simultaneous quantitation of 10 transplant-related double-stranded DNA (dsDNA) viruses (adenovirus [ADV], BK virus [BKV], cytomegalovirus [CMV], Epstein-Barr virus [EBV], human herpesvirus 6A [HHV-6A], HHV-6B, herpes simplex virus 1 [HSV-1], HSV-2, JC virus [JCV], and varicella-zoster virus [VZV]). The mNGS 95% limit of detection ranged from 14 copies/ml (HHV-6) to 191 copies/ml (BKV), and the lower limit of quantitation ranged from 442 international units (IU)/ml (EBV) to 661 copies/ml (VZV). An evaluation of 50 residual plasma samples with at least one DNA virus detected in prior clinical testing showed a total percent agreement of mNGS and quantitative PCR (qPCR) of 89.2% (306/343), with a κ statistic of 0.725. The positive percent agreement was 84.9% (73/86), and the negative percent agreement was 90.7% (233/257). Furthermore, mNGS detected seven subsequently confirmed coinfections that were not initially requested by qPCR. Passing-Bablok regression revealed a regression line of  = 0.953 + 0.075 (95% confidence interval [CI] of the slope, 0.883 to 1.011; intercept, -0.100 to 0.299), and Bland-Altman analysis (mNGS - qPCR) showed a slight positive bias (0.28 log concentration; 95% limits of agreement, -0.62 to 1.18). In conclusion, the mNGS-based Galileo pipeline demonstrates analytical and clinical performance comparable to that of qPCR for transplant-related DNA viruses.

BT - J Clin Microbiol DA - 2019 12 DO - 10.1128/JCM.01113-19 IS - 12 J2 - J Clin Microbiol LA - eng N2 -

Infections with DNA viruses are frequent causes of morbidity and mortality in transplant recipients. This study describes the analytical and clinical performance characteristics of the Arc Bio Galileo Pathogen Solution, an all-inclusive metagenomic next-generation sequencing (mNGS) reagent and bioinformatics pipeline that allows the simultaneous quantitation of 10 transplant-related double-stranded DNA (dsDNA) viruses (adenovirus [ADV], BK virus [BKV], cytomegalovirus [CMV], Epstein-Barr virus [EBV], human herpesvirus 6A [HHV-6A], HHV-6B, herpes simplex virus 1 [HSV-1], HSV-2, JC virus [JCV], and varicella-zoster virus [VZV]). The mNGS 95% limit of detection ranged from 14 copies/ml (HHV-6) to 191 copies/ml (BKV), and the lower limit of quantitation ranged from 442 international units (IU)/ml (EBV) to 661 copies/ml (VZV). An evaluation of 50 residual plasma samples with at least one DNA virus detected in prior clinical testing showed a total percent agreement of mNGS and quantitative PCR (qPCR) of 89.2% (306/343), with a κ statistic of 0.725. The positive percent agreement was 84.9% (73/86), and the negative percent agreement was 90.7% (233/257). Furthermore, mNGS detected seven subsequently confirmed coinfections that were not initially requested by qPCR. Passing-Bablok regression revealed a regression line of  = 0.953 + 0.075 (95% confidence interval [CI] of the slope, 0.883 to 1.011; intercept, -0.100 to 0.299), and Bland-Altman analysis (mNGS - qPCR) showed a slight positive bias (0.28 log concentration; 95% limits of agreement, -0.62 to 1.18). In conclusion, the mNGS-based Galileo pipeline demonstrates analytical and clinical performance comparable to that of qPCR for transplant-related DNA viruses.

PY - 2019 T2 - J Clin Microbiol TI - Metagenomic Next-Generation Sequencing for Identification and Quantitation of Transplant-Related DNA Viruses. VL - 57 SN - 1098-660X ER -